Interacoustics Aqua Stim Manual De Instrucciones página 26

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 96
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The AquaStim is intended for use in the electromagnetic environment specified below. The customer or
the user of the AquaStim should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test level
Electrostatic
+8 kV contact
Discharge (ESD)
+15 kV air
IEC 61000-4-2
+2 kV for power supply
Electrical fast
lines
transient/burst
+1 kV for input/output
IEC61000-4-4
lines
Surge
+1 kV differential mode
IEC 61000-4-5
+2 kV common mode
< 5% UT (>95% dip in
UT)
for 0. 5 cycle
Voltage dips, short
40% UT (60% dip in
interruptions and
UT)
voltage variations
for 5 cycles
on power supply
lines
70% UT (30% dip in
UT)
IEC 61000-4-11
for 25 cycles
<5% UT (>95% dip in
UT)
for 5 sec
Power frequency
(50/60 Hz)
3 A/m
IEC 61000-4-8
Note: UT is the A. C. mains voltage prior to application of the test level.
Aqua Stim Water Irrigator – Instructions for Use - EN
Compliance
+8 kV contact
+15 kV air
+2 kV for power
supply lines
+1 kV for input/output
lines
+1 kV differential
mode
+2 kV common mode
< 5% UT (>95% dip
in UT)
for 0. 5 cycle
40% UT (60% dip in
UT)
for 5 cycles
70% UT (30% dip in
UT)
for 25 cycles
<5% UT
for 5 sec
3 A/m
Electromagnetic environment -
guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
greater than 30%.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment.
Mains power quality should be
that of a typical commercial or
residential environment. If the
user of the AquaStim requires
continued operation during power
mains interruptions, it is
recommended that the AquaStim
be powered from an
uninterruptable power supply or
its battery.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or residential
environment.
Page 20