Veridian Healthcare 11-800 Manual De Instrucciones página 15

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

ELECTROMAGNETIC COMPATABILITY
for me eQuIpment and me systems that are not lIfe-supportIng
guidance and manufacturer's declaration - electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below.
The customer or the user of this device should assure that it is used in such an environment.
Immunity Test
Radiated RF
3 V/m 80 MHz to
IEC 61000-4-3
2,5 GHz
note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
note 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
* Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which this
monitor is used exceeds the applicable RF compliance level above, this monitor should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-
orienting or relocating this monitor.
** Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3V/m.
IEC 60601
Compliance
Test Level
Level
3 V/m
Table 3
Electromagnetic Environment Guidance
Portable and mobile RF communications equipment should be
used no closer to any part of this monitor, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
recommendedseparationdistance
d =
1
2 .
d =
2 .
1
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d = 2.3
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance
in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,* should be less than the
compliance level in each frequency range.**
Interference may occur in the vicinity of equipment marked
with the following symbol:
15
English •

Capítulos

Tabla de contenido
loading

Tabla de contenido