ELECTROMAGNETIC COMPATABILITY

guidance and manufacture's declaration - electromagnetic emissions
This device is intended for use in the electromagnetic environment specified below.
The customer or the user of this device should assure that it is used in such an environment.
Emission Test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/flicker
emissions IEC 61000-3-3
guidance and manufacture's declaration - electromagnetic immunity
This device is intended for use in the electromagnetic environment specified below.
The customer or the user of this device should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level Compliance Level
Electrostatic
± 6 kV contact
discharge (ESD)
± 8 kV air
IEC 61000-4-2
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
14
• English
for all me eQuIpment and me systems
Compliance
Group 1
Class B
Not applicable
Not applicable
for all me eQuIpment and me systems
± 6 kV contact
± 8 kV air
3 A/m
Table 1
Electromagnetic Environment Guidance
This device uses RF energy only for its internal
function. Therefore, its RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
This device is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used for
domestic purposes.
Table 2
Electromagnetic Environment Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.