Compact Digital Multimeter / English
This symbol adjacent to one or more terminals identifies them as
being associated with ranges that may, in normal use, be subjected
to particularly hazardous voltages. For maximum safety, the meter
and its test leads should not be handled when these terminals are
energized.
3. Controls and jacks
1. LCD Display
2. Data Hold Button
3. Function Switch
4. Backlight Button
5. 10 A Jack
6. COM Jack
7. Positive Jack
Note: Tilt stand, fuse and battery
compartment are on rear of unit.
4. Symbols and Annunciators
Continuity
Diode test
μ
μ Micro(amps)
M
M Milli(volts, amps)
K
K Kilo(ohms)
Ω
Ω O hms
VDC
VDC Volts direct current
VAC
VAC Volts alternating current
ADC
ADC Amps direct current
BAT
BAT Battery test
2
5. Specifications
Function
DC Voltage
(V DC)
AC Voltage
(V AC)
DC Current
(A DC)
Resistance
Battery Test
NOTE: Accuracy specifications consist of two elements:
NOTE: Accuracy is stated at 65° F to 83° F (18°C to 28°C) and less than 75% RH.
02/01/2018 Version No. 001
02/01/2018 Version No. 001
Compact Digital Multimeter / English
Range
Resolution
Accuracy
200 mV
0.1 mV
2000 mV
1m V
±(0.5%reading + 3digits)
20 V
0.01 V
200 V
0.1 V
±(0.8%reading + 3digits)
600 V
1 V
200 V
0.1 V
±(1.2%reading + 10digits 50/60 Hz)
600 V
1 V
200 μA
0.1 μA
±(1.0%reading + 5digits)
20 mA
10 μA
200 mA
100 μA
±(1.2%reading + 5digits)
10 A
10 mA
±(3.0%reading + 5digits)
200 Ω
0.1 Ω
2 kΩ
1Ω
±(1.0%reading + 5digits)
20 kΩ
0.01 kΩ
200 kΩ
0.1 kΩ
2 MΩ
1kΩ
±(1.5%reading + 5digits)
9 V
10 mV
±(1.0%reading + 2digits)
1.5 V
10 mV
• (% reading) – This is the accuracy of the measurement circuit.
• (+ digits) – This is the accuracy of the analog to digital converter.
3