Troubleshooting Table; Appendix: Serial Connection Protocol - Selecta S-2000 Manual De Instrucciones

Analizador clinico semiautomatico
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL CODE 80202 REV A
J
.P. SELECTA s.a.u.
Feb/2012

13.7 Troubleshooting table

4) Non aspiration of sample
5) The result of water blank is too high
6) Repeats are not good
7) Controls out of the target
8) The aspirate volume not stable

Appendix: Serial Connection protocol

1. Serial port state:
Data is transmitted at 19200 baud,
With 1 stop bit
8 data bit and Even Parity
2. Message format
The initial parameter must be the capital letter "R", each line consists of 12 data
in following order:
Parameter
Initial parameter
Patient name
Department
Sex
Patient No.
Name of test
Result
Units of measurement
Checksum
Checksum = All of parameter + All of space mark
3. Send Over
S-2000 sends a character string "TRANSFER FINISH", when the task has been com-
pleted.
Example:
R,Biachi,, 0, 43, 1, TP, 136.00, g/L, 0794
R,Rossi,,1, 40, 2, TP, 147.00, g/L, 0765
R,Tony,,0,6,3,TP,35.00,g/L,069d
TRANSFER FINISH
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: http://www.jpselecta.es
(It can be modified without notice)
— Check the pump tube
— Check tube pipeline
— Rinse and check the flow cell
— Check the lamp
— Change the distilled water
— Rinse the flow cell
— Check if there is air bubble in flow cell
— Reagent validity
— Assure the program is correct.
— The quality of controls
— To test again by changing the mode
— Check the flow cell and select a new
reagent and control
— Liquid system blocked, Rinse the tube
pipeline, or change the tube.
Possible values Space mark
"R"
Character or NULL
Character or NULL
Character or NULL
"0" = Male
"1" = Female
Character
Character
Character
Character
Character