INSTRUCTION MANUAL CODE 80202 REV A
Screen 6.6 System settings menu
Screen 6.7 Fixed Test menu
Screen 6.8 Test Enter menu
Screen 6.9 Programming menu
Screen 6.10 Calibration menu.
J
.P. SELECTA s.a.u.
Feb/2012
System settings menu:
1. Setting time and date
2. Select pump speed (Factory set at 6)
3. Adjusting contrast of LCD
4. Printer Select
Printer Inside: Thermal printer is selected.
Printer Outside: External printer.
5. Enter the Hospital or laboratory name
6.4 Programming test parameters
S-2000 Analyser has a built-in menu of 130 assays. 47 assays of most commonly
applied data reduction formulae are preprogrammed.
6.4.1 Modifying stored programs
Click "Test" button in main menu. The screen 6.7 will be displayed.
1. Click the button which will be modified, such as "UA", The screen 6.8 will ap-
pears.
Select "Edit" to modify the parameter of the test. Scree 6.9.
2. Modifying parameters according to the reagent instruction.
3. Click "Save" button to save the modification, or click "Cancel" to ignore.
6.4.2 Programming a new assays
1. Select a free button of assays by clicking "Pgdn" button, then click the button.
2. Input parameters
3. Save the input
6.4.3 Calibration parameter
Select "Calibration", "Calibration setting" window will be open.
Screen 2.10 Calibration setting menu
1) Calibration for linear tests
•
Standard number means that how many standards will be measured.
•
The measured results will be averaged. Input the number by clicking the window.
•
Repeat number: This parameter determines how many standards (with the same
concentration) will be used for calibration. The results of multiple measurements
will be averaged by the system.
•
Select "OK" to confirm the selection.
•
Select "Exit" to go back to above menu with modify
2) Calibration for non linear tests
•
Standard number means that how many standards will be measured.
•
The standards are different concentrations. The measured results will be a calibra-
tion curve. Input the number by clicking the window.
•
Repeat number: This parameter determines how many standards (with the same
concentration) will be used for calibration. Possible inputs are 1, 2 and 3. The
results of multiple measurements will be averaged by the system.
•
Select "OK" to confirm the selecting.
•
Select "Exit" to go back to above menu.