TECHNICAL SPECIFICATIONS

Insulation:..
Overvoltage.category:... CATIV.600V
Note:
These.meters.meet.CAT.III.and.CAT.IV.IEC.61010.standards..The.IEC.61010.safety.standard.
defines four overvoltage categories (CAT I to IV) based on the magnitude of danger from transient
impulses. CAT III meters are designed to protect against transients in fixed-equipment installations
at the distribution level; CAT IV meters are designed to protect against transients from the primary
supply.level.(overhead.or.underground.utility.service).
Maximun.voltage.between.any.terminal.and.earth.ground:..600V.DC/AC.RMS
Display:..
Polarity:..
Over-range:..
Low.battery.
indication:.
Measurement.rate:..
Auto.power.off:..
Operating.environment:...-10.°C.to.50.°C.(14.°F.to.122.°F).at.<.70.%.relative.humidity.
Storage.temperature:.
Relative humidity:
Class2,.Double.insulation..
11000.counts.LCD.display
Automatic,.(-).negative.polarity.indication
"OL".mark.indication.
A. battery. symbol. is. displayed. when. the. battery. voltage. drops. .
below.the.operating.level.........
4.times.per.second.nominal.
Meter.automatically.shuts.down.after.approx..15.minutes.of.inactivity..
-30.°C.to.60.°C.(-4.°F.to.140.°F).at.<.80.%.relative.humidity.to
90% (0°C to 30°C); 75%(30°C to 40°C); 45%(40°C to 50°C);