Selecta pH-2003 Manual De Instrucciones página 9

Tabla de contenido
Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL CODE 80193 REV A
2
8
3
4
5
6
J
.P. SELECTA s.a.
5 Description
1
All unit functions are selected from the front keypad. The displays shows the pH
reading, the temperature set and the pH-meter status.
1. Electrode BNC connector.
2. Display.
3. On/Off.
4. Push button Mode
5. Push button Enter
9
6. Push button Up.
7. Push button Down.
8. Push button display back light
9. Electrode holder
7
6 Operation
6.1 Placing batteries
Place the batteries on the rear side holder.
Check correct polarity.
If the pH-meter will no be used for a long time remove the batteries.
When message «LoBat» appears on the display, change the batteries.
6.2 Electrode connection
The electrode must be stored with its immersion cap fi lled with electrolyte.
Transport the pH-meter in the briefcase with the electrode unplugged, and the pH-
meter protected with the short-circuit connector.
To connect the electrode, remove the short-circuit connector from the pH-meter
and to connect the electrode.
Note: The BNC connectors are extracted and placed turning 1/4 turn.
Keep the electrode with its immersion cap until use.
The immersion cap contend KCl 1M as electrolyte which is cor-
rosive.
Take suitable cautions to avoid electrolyte be in contact with
skin or wear. Avoid any contact with eyes.
6.3 Keypad
Press Start/Stop to switch on the pH-meter.
Press, successively, «Mode» to select the different operation:
«ºC»:
«STD1»:
«STD2»:
«MEAS»:
Ctra. NII Km 585.1 Abrera 08630 (Barcelona) España
e-mail: [email protected] - website: www.grupo-selecta.com
APRIL 2013
(It can be modifi ed without notice)
CAUTION:
To enter the measuring temperature.
Input standard 1.
Input standard 2.
Selection between «pH» or «mV».
Page: 9
Tel 34 937 700 877 Fax 34 937 702 362