How to Disinfect

The Micro-Air
®
nebulizer can be cleaned and disinfected easily because the Medication
Bottle and the Mesh Cap can be removed from the Main Unit. The unit can
be disinfected with benzalkonium chloride.
1
Make 0.1% dilute solution using benzalkonium chloride, and soak the
Medication Bottle and Mesh Cap in the solution for 10 minutes.
2
Take the Medication Bottle and Mesh Cap out of solution
to rinse it with water.
3
Wipe off the remaining water for faster drying.
4
If using commercial cleaners, follow manufacturers instructions.
Caution
Do not use disinfection solution containing sodium hypochlorite.
• The unit may not be able to be used after getting covered with rust.
5
The following parts can be disinfected:
Medication Bottle
Mesh Cap
Mask
Mouthpiece
Mouthpiece Adapter
18

How to Replace the Mesh Cap

You can usually use the Mesh Cap for approximately 1 year (according to the test results of OMRON).
However, this may vary depending on medication type and individual use and care. In some cases,
the unit may not nebulize or may nebulize weakly even after disinfection is done.
If the unit does not nebulize or nebulizes poorly even after cleaning the Mesh Cap,
purchase a new Mesh Cap from your nearest Omron dealer or where you purchased your unit from.
The part number is U22-4.
1
Open the Mesh Cap.
Lift up the Mesh Cap
by strongly applying
your thumb on the brim.
2
Remove the Mesh Cap from the
Medication Bottle.
3
Install a new Mesh Cap.
Insert a new
Mesh Cap
vertically.
Lift up the
Mesh Cap
vertically.
Close it
securely.
19