SP-04-2013 OK (830P.) EN-DE FR-ES-PL-RUS_830P. 13-04-13 15:11 Strona 10
WARNING!
If disposable batteries are used, it is unacceptable to connect the device to the charger.
In order to prolong the life span of the rechargeable batteries, the following rules should be
followed:
1. Before the pipette controller is activated for the first time, the batteries should be charged.
2. Avoid short charging cycles – disconnect the stand from power supply during work
when charging is not needed.
3. If the pipette controller starts to indicate low battery level during work, connect it to the
power supply to continue working.
4. Do not leave the pipette controller discharged for a long period of time.
5. In case of a planned longer break in the work of the device, it is recommended to take
the batteries out from the battery compartment.
9 - MAINTENANCE
Cleaning
The SWIFTPET PRO does not require any maintenance. Its external parts may be cleaned
with a swab moistened with isopropyl alcohol.
The nosepiece and the pipette holder may be autoclaved at 121 °C for 20 mi nutes. After auto-
claving, dry the pipette holder. The filter included in the set may be sterilised by autoclaving
at 121°C for not more than 15 minutes.
Storage
The SWIFTPET PRO shall be stored in a dry place. The allowable storage temperature: -20 °C
to +50 °C.
During breaks in the work the pipette controller shall be placed on a charging stand in hori-
zontal position (fig. 7A) or in vertical position (fig. 7B). The charging stand may be fitted on
a wall (fig. 7B) with a fixing set included with the pipette controller. If the pipette controller is
stored on a charging stand (fig. 7A or fig. 7B), the charging stand should be disconnected
from the power supply.
CAUTION:
Do not store the pipette controller with a filled pipette.
10 - COMPONENTS
The SWIFTPET PRO set is supplied with the following components:
• Pipette controller
• Charger
• Charging stand
• Membrane filter 0.2 μm
• Batteries – 3 pcs.
• Short instruction manual
• CD