Chlorine Hr With Tablet - Lovibond Tintometer MD100 Manual De Instrucciones

Ocultar thumbs Ver también para Tintometer MD100:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 52
Methods
GB
CLHr
0.0.0
Zero
Test
CLHr
RESULT
Notes:
1. Oxidising agents interfere as they react in the same way as Chlorine.
Reagent
Set ACIDIFYING GP/
CHLORINE HR (KI)
CHLORINE HR (KI)
ACIDIFYING GP
MD100_7c 03/2020

Chlorine HR with Tablet

5 – 200 mg/l
Insert the adapter for 16 mm vials.
Fill a clean vial (16 mm Ø) with 8 ml of the water sample and perform
zero calibration (see "Operation").
Add one CHLORINE HR (KI) tablet straight from the foil to the water
sample and crush the tablet using a clean stirring rod.
Add one ACIDIFYING GP tablet straight from the foil to the same
water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablets are dissolved.
Place the vial in the adapter making sure that the ı marks are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l Chlorine.
Tolerance: ± 5 mg/l
Form of reagent/Quantity
Tablet / per 100
inclusive stirring rod
Tablet / 100
Tablet / 100
Order-No.
517721BT
513000BT
515480BT
21
Tabla de contenido
loading

Este manual también es adecuado para:

Md100

Tabla de contenido