Error messages
Error
NO TUNING
TEST SETTINGS
SECURITY PLATE
HUMIDITY
CU NO REACTION
CU TIMEOUT
PROCESS-TIMELIMIT
POS. ERROR
SIGMATAX SIGNAL
SIGMATAX OFF
LEVEL LIMITS
LEVEL OFFSET
TEMPERATURE SENSOR
AMPLIFIER
CALIBRATION FACTOR
INTERNAL BUS ERROR
Cause
Electronic error
Possible device adjustment
Safety panel loose
Moisture sensor responds
Electronic error
Undefined piston valve position
SIGMATAX error
Switched off SIGMATAX
Photometer error
Temperature measurement defective
Photometer error
Electronic error
Measures
Call customer service
Call up settings and check completely, with frequent occurrence
inform customer service
Check attachment of the safety panel
Remedy cause (leak), dry off moisture sensor and acknowledge error in [Status]
menu
Call customer service
Check piston valve and acknowledge error in [STATUS] menu
Call up [STATUS] menu and determine error cause
Check fiber optic and SIGMATAX
After acknowledgement in [STATUS] menu and another occurrence call customer
service
Call customer service
Check of the dosing and of the reagents
Call customer service
English 111