C2.4 Closing the Chamber Door

 The door will automatically lock at the start of a high level disinfection cycle.
 If the door is not properly closed, screen message: Close Chamber Door will be displayed.

C2.5 Disinfecting the Probe

The next screen message: Is the Probe Clean and Dry?
1. If the probe has been pre-cleaned and dried, press Yes using the soft key button. If Yes is
selected, you will be prompted by the screen message: Press start to begin.
2. If the probe has NOT been pre-cleaned and dried, press No. Remove the probe, clean and dry
the probe and position again as directed in section C2.3.
3. Once the probe is pre-cleaned and dried, select Yes and press the START button to initiate the
cycle or Cancel to unlock the machine and remove the probe.
4. The progress of the disinfection cycle is indicated on the LCD screen message: Disinfecting.
5. The High Level Disinfection cycle will take 7 minutes to complete.
WARNING: In the unlikely event mist is visibly escaping from the chamber, remain at a
distance from the trophon until completion of the operating cycle and the mist is no
longer visible. Do not come into direct contact with the mist. Contact your customer
service representative (see Part D – Troubleshooting).

C2.6 Removing the Probe

1. When the cycle has been completed, the trophon will sound an audible alarm.
NOTE: After completion of a successful high level disinfection cycle, the ultrasound probe and
chamber may have surface temperatures up to 45
L00049 3.0 04/2016
Figure 12
o
o
o
C/ 113
F and 60
C/ 140
o
F respectively. Care
13