Technical Specifications
Operating pressure range
Maximum single fault pressure
Pressure measurement
tolerance
Flow measurement tolerance
VAuto mode
S, ST and T modes
CPAP mode
Sound pressure level
Dimensions (L x W x H)
Weight
Power supply
Housing construction
Operating temperature
Operating humidity
Storage and transport
temperature
Storage and transport humidity
Operating altitude
Electromagnetic compatibility
Air filter
Air tubing
Air outlet
IEC 60601-1 classification
Note: The manufacturer reserves the right to change these specifications without notice.
Symbols Which Appear On The Device
Attention, consult accompanying documents;
Dangerous voltage;
14
3 to 25 cm H
O
2
40 cm H
O
2
±0.5 cm H
O ± 4% of the measured reading
2
±0.1 or 20% of reading, whichever is greater
4 to 25 cm H
O (measured at the mask); Min EPAP: 4 cm H
2
Max IPAP: 25 cm H
O; Pressure Support: 0 to 10 cm H
2
IPAP: 4 to 25 cm H
O (measured at the mask); EPAP: 3 to 25 cm H
2
(measured at the mask); Pressure Support: 0 to 22 cm H
4 to 20 cm H
O (measured at the mask)
2
26±2 dBA as measured according to ISO 17510-1: 2002
4.4" x 6.5" x 5.7" (112 mm x 164 mm x 145 mm)
2.9 lb (1.3 kg)
Input range 100–240V, 50–60Hz, 40VA (typical power consumption),
< 100VA (maximum power consumption)
Flame retardant engineering thermoplastic
41 º F to 95 º F (+5 º C to +35 º C)
10–95% non-condensing
-4 º F to 140 º F (-20 º C to +60 º C)
10–95% non-condensing
Sea level to 8,500' (2,600 m)
Product complies with all applicable electromagnetic compatibility
requirements (EMC) according to IEC60601-1-2, for residential,
commercial, and light industry environments
Two-layered, powder-bonded, polyester non-woven fiber
Flexible plastic, 1 x 6'6" (2 m)
The 22 mm conical air outlet complies with ISO 5356-1
Class II (double insulation), Type CF
Class II equipment;
Drip proof;
Type CF equipment;
Start/Stop.
O;
2
O
2
O
2
O
2