Service And Maintenance - MSA AUER SSR 30/100 Manual Del Usuario

Autorescador de oxígeno
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 19
inhalt 06/03
13.06.2003 11:02 Uhr
7.2

Service and Maintenance

1)
Intervals
Measures
Before and after
Inspection
each carrying
Annually
2)
After 3 years
Inspection
2)
After 5 years
Full inspection
2)
After 8 years
Inspection
After 10 years
End of service life.
Defective apparatus must not be used, must be
removed from service and returned to MSA for
inspection.
User specific conditions may require shorter
intervals.
The service life depends on the type and kind of
treatment to which the apparatus were subjected.
1)
The intervals are based on the date of manufac-
ture (month/year) on the container bottom.
Apparatus manufactured in the same year may
be combined for inspections.
2)
A service contract with the MSA Customer
Service is recommended
3)
Complete-check = includes full internal exami-
nation. The result determines about the further
use of apparatus.
4)
Tightness check with leak tester D1118845 or by
MSA service.
5)
Check if the lead seal is intact and no outer
damages exist.
7.3
Records
Records should be kept for each SSR. These
include serial number, date of manufacture date
apparatus was put into service as well as mainte-
nance measures.
6
Seite 6
8
Training with the SSR 301100 TR
In order to assure that donning and operating the
apparatus in an emergency is done propperly and
without delay, the training apparatus SSR 30/100
External visual
TR is available. With it handling and breathing of
5)
check
the apparatus can be practised without using a
chemical canister. Such training should be re-
4)
Tightness check
peated at regular intervals. The training apparatus
is comparable to the emergency apparatus as far
Tightness check
as size, weight, outer design, handling and breath-
and complete-
ing resistance are concerned.
check of all
apparatus showing
However, the breathable air is taken from the
significant signs
ambient atmosphere, i. e. the exercises must be
3)
of wear and tear
performed in contaminant-free air. To distinguish it
from the emergency apparatus the training
Complete-check
apparatus is marked blue.
3)
of all apparatus
Tightness check
and complete-
check of all
apparatus showing
9
Apparatus and Accessories
significant signs
3)
of wear and tear
No. Description
1
Oxygen Self-Rescuer
MSA AUER SSR 30/100
2
Training Self-Rescuer
MSA AUER SSR 30/100 TR
3
Reconditioned apparatus
MSA AUER SSR 30/100
4
Tightness tester for container
with apparatus
5
Insert SSR 30/100 for
tightness tester
6
Belt
Part No.
D1123 701
D1123 711
D1123 703
D1118 845
D1118 322
D1113 958
Tabla de contenido