CALIBRATING THE BOTTLE SCALE

Disconnect the machine from the electrical supply.
between 55 and 72lbs (28 to 32 kg). Remove the plastic cover from the rear of the machine to
access the machine bottle. Close the blue and red taps on the bottle. Unscrew the bottle lock
nut (3 – Fig. 4) Separate the heating coil (r – Fig. 8) from the bottle being careful not to touch or
disconnect the heater wires.. Remove the bottle (d – Fig. 8) from its lodging leaving the heater
around the scale plate. Set the bottle on a stand at least 15 inches (40 cm) in height.
Access the CALIBRATION MENU as explained above.
The following screen will be displayed:
With BOTTLE SCALE flashing, press ENTER. The following screen will be displayed:
Press the ENTER key.
With the bottle lifted off the scale plate, press START. The following screen will be displayed:
INSTRUCTIONS MANUAL
AUTOMATIC PROCEDURE
ASSISTED PROCEDURE
<NEXT MENU>
Gas avail
X,XX G
<CALIBRATION>
DATA AND CONFIGURAT.
SERVICE ALARM
BOTTLE FILLING
To enter calibration
insert code
....
And press ENTER
<BOTTLE SCALE >
BOTTLE PRESSURE TR.
BOTTLE TEMPERATURE
PREVIOUS MENU
<CALIBRATION>
BOTTLE DATA
PREVIOUS MENU
ZERO LEVEL
Raise the bottle
Levels
xxxxx
press START
-18-
Procure a known reference weight