ResMed AirMini Manual Del Usuario página 24

Ocultar thumbs Ver también para AirMini:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Maximum single fault steady pressure
Machine will shut down in the presence of a single fault if the steady state pressure
exceeds 30 cm H
1 sec.
Sound
Declared dual-number noise emission values in accordance with ISO 4871:1996
Pressure level measured according to ISO
80601-2-70:2015 (CPAP mode)
Power level measured according to ISO 80601-
2-70:2015 (CPAP mode)
The A-weighted sound pressure level measured according to ISO 17510:2015:
Air tubing connected to AirMini F20 connector
with mask
Air tubing connected to AirMini N20 or P10
connectors with mask
The A-weighted sound power level measured according to ISO 17510:2015:
Air tubing connected to AirMini F20 connector
with mask
Air tubing connected to the AirMini N20 or P10
connectors with mask
AirMini machine
Dimensions:
Weight:
Housing construction:
Air outlet:
Air filter
Standard:
22
O (30 hPa) for more than 6 sec or 40 cm H
2
O (40 hPa) for more than
2
30 dBA with uncertainty of 2 dBA
38 dBA with uncertainty of 2 dBA
19 dBA with uncertainty of 3 dBA
17 dBA with uncertainty of 3 dBA
27 dBA with uncertainty of 3 dBA
25 dBA with uncertainty of 3 dBA
5.4" (W) x 3.3" (D) x 2.1"(H)
136 mm (W) x 84 mm (D) x 52 mm (H)
300g (10.6oz)
Flame retardant engineering thermoplastic
Proprietry connector, 16mm inner
diameter. Not compatible with
EN ISO 5356-1 connectors.
Material: Polyester non woven fiber
Average arrestance: >75% for ~7 micron
dust