If power is interrupted during therapy, the machine will automatically restart therapy
when power is restored.
AirMini power
Standby power consumption:
Environmental conditions
Operating temperature:
Operating humidity:
Operating altitude:
Storage and transport temperature:
Storage and transport humidity:
Electromagnetic compatibility
The AirMini complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2:2014, for residential, commercial and light industry
environments. Portable and mobile RF communications equipment should be used no
closer to any part of the machine, including cables, than the recommended 10 cm
separation distance. The AirMini has been designed to meet EMC standards. However,
should you suspect that the machine performance (eg, pressure or flow) is affected by
other equipment, move the machine away from the possible cause of interference.
The AirMini complies with Part 15 of the FCC Rules and Industry Canada license-exempt
RSS standards. Operation is subject to the following two conditions: This machine may
not cause harmful interference, and this machine must accept any interference received,
including interference that may cause undesired operation.
FCC ID: QOQBT121, IC: 5123A-BGTBT121
Additional information regarding the FCC Rules and IC compliance for this machine can
be found on www.resmed.com/downloads/devices.
IEC 60601-1 (Edition 3.1) classification
Class II (double insulation), Type BF, Ingress protection IP22.
1W
+41°F to +95°F (+5°C to +35°C)
Note: Under extreme ambient temperature
conditions (104°F/40°C) the air and parts of
the tubing near the mask may reach
temperatures up to 109°F/43°C. These parts
of the tubing are not in contact with the user
and the machine remains safe under these
extreme conditions.
10 to 95% relative humidity, non-condensing
Sea level to 8,500' (2,591 m); air pressure
range 1013 hPa to 738 hPa
-13°F to + 158°F (-25°C to + 70°C)
5 to 95% relative humidity, non-condensing
English
21