Operation:
Storage:
Transport:
Battery performance:
Battery over current/temperature
protection:
Light emitting diode:
Average light intensity:
Output peak wavelength range:
Unit handpiece dimensions:
Unit weight:
Complies with the following Directives/Standards:
93/42/EEC
2002/95/EC
IEC 60601-1
IEC 60601-1-2
ISO 10650
EN 62471
IEC 62471
Council Directive 93/42/EEC of 14 June 1993 concerning medical de-
vices amended by directive 2007/47/EC, annex 1
Restriction of the use of hazardous substances in electrical and elec-
tronic equipment
2005 – Medical Electrical Equipment (General requirements for basic
safety and essential performance)
2005 – Medical electrical equipment – Part 1-2: General requirements
for safety and essential performance – Collateral standard: Electromag-
netic compatibility – Requirements and tests
2015 – Dentistry – Powered polymerization activators
2008 – Photobiological Safety of Lamps and Lamp Systems
2006 – Photobiological Safety of Lamps and Lamp Systems
Ambient temperature:
Between 0 °C and +30 °C (32 °F and 86 °F)
Relative humidity: Between 20% and 90%
Ambient temperature:
Between -10 °C and +35 °C (14 °F and 95 °F)
Ambient temperature:
Between -10 °C and +50 °C (14 °F and 122 °F)
•
Low self-discharge technology results in a long
operating life.
•
Battery is pre-charged and ready to use upon
purchase.
•
Time for battery recharge: Approximately
3 hours.
•
Automatic Quick Charging Mode: allows a
minimum of 5 minutes operating time after
10 minutes recharging.
Resettable fuse
3 W LED
Approximately 1000 mW/cm
Between 460 nm and 490 nm
(intensity maximum around 475 nm)
Length: 240 mm
Width: 31 mm
Handpiece with battery pack: 90 grams
Charging base with power connector: 200 grams
(CureRite
™
)
2
11