Dynamic Short-Term Pressure Accuracy with and without
DV5HH Humidifier
Waveform
Sinusoidal cycle with
I:E = 1:1
Equipment classification with respect to protection from electric shock . . . . . . . . . . .Class II
Degree of protection from electric shock . . . . . . . . . . . . . . . . . . . . . . . Type BF Applied Part
Degree of protection against ingress of liquids . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IPX1
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or
with oxygen or nitrous oxide.
Mode of operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Continuous
Clinical Specifications for Digital Outputs
Mask Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 25.5 cmH2O ± 0.5
Patient Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . -127 to +127 L/min ± 10 L/min
Leak Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to +127 L/min ± 10 L/min
Estimated Tidal Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 1023 mL
NOTe–Tidal Volume is an estimate. It is provided for trending purposes only.
SE-DV54-1
30
(per ISO 17510-1:2007)
Volume
Breath Rate
[mL]
[min-1]
500
10, 15, 20
SpEcificationS
Set Pressure
Pressure Accuracy
[cmH2O]
[Pk-Pk cmH2O]
3.0
7.0
12.0
16.0
20.0
0.5
0.5
1.0
1.0
1.0