Record Paper - KONTROLab EC-300 Manual De Usuario

Electrocardiógrafo veterinario
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

3) Replacement
When the useful life of battery is over, or foul smell and leakage has been found, please
contact with manufacturer or local distributor for replacement of battery.
WARNING
Only qualified service engineer authorized by EDAN can open the battery
compartment and replace the battery. And the battery of same model and
specification provided by manufacturer must be used.
Danger of explosion -- Do not reverse the anode and cathode when
connecting the battery.
When the battery's useful life is over, contact with the manufacturer or local
distributor for disposal or dispose the battery according to local regulations.

8.3.2 Record Paper

Storage requirements:
Record paper should be stored in dry, dark and cool area, avoiding excessive
temperature, humidity and sunshine.
Do not put the paper under fluorescence for long time.
Be sure that there is no polyvinyl chloride or other chemicals in the storage
environment, which will lead to color change of the paper.
Do not overlap the recorded paper long time, or else the ECG record may trans-print
each other.
Note: Record paper provided by manufacturer should be used. Other paper may
shorten thermal print head's life. And the deteriorated print head may lead to
illegible ECG record and block the advance of paper etc.
:
36
Tabla de contenido
loading

Tabla de contenido