Chapter 3 Measurement - DRAMINSKI GMM Manuel De Instrucciones

Ocultar thumbs Ver también para GMM:
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 101
Having performed these actions you can put the
dosage tube on the measurement chamber of the
meter (on the lower edge of the feeder there is a
special slant). You should then pull out the slider,
which will cause the sample to fall into the meas-
urement chamber.
ATTENTION!
While pouring the seeds you should pull
out the slider with a steady, but decided
movement which lasts about 1 second.
Upper edges of the measurement chamber
(red elements) should be completely cov-
ered by seeds after filling.
e) Having filled the measurement chamber you
should not move or shake the device. You
should only press the button "OK".
Under the name of the sample there will be the
information: "analysing" displayed. After a
short time the result of the measurement will
be shown – the percentage of humidity and
temperature in C degrees. (the final humidity
reading will appear after the temperature's in-
fluence is taken into account).
f) After measurement you should empty the
chamber and press "OK" button.
The command "pour in seed" will be displayed.
It means, you can repeat the measurement for
the same sample or check the humidity for an-
other sample (of the same species).
g) To choose another species (another name of
the sample) you should press the button
"CAL/
".
CANCEL
h) After finishing measurements the tester should
be turned off by pressing "ON/OFF" button.
If the tester is not being used, it will turn off au-
tomatically after 3 min.
ATTENTION!
Please remember, to pour seeds into the
measurement chamber only after the com-
mand "pour in seeds" is displayed.
CHAPTER 3
MEASUREMENT
41