VATECH PaX-i3D Smart Manual Del Usuario página 29

Ocultar thumbs Ver también para PaX-i3D Smart:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 236
3 Acquiring PANO Images
3.2.2
TMJ bite
Sinus Chinrest
28
Completing Patient Positioning
1. After confirming patient positioning and beam alignment, adjust the temple
supports to fit snugly on both sides of the patient's head by clicking the temple
support adjustment button.
Make sure that the Temple Support is in Close position
before clicking the READY button.
CAUTION
2. Click the Ready button after the patient has been properly positioned. No
X-Ray will be emitted at this point.
3. Proceed to section 3.3 Initiating X-Ray Exposure.
TMJ Open
There are two TMJ sub modes: TMJ Open and TMJ Close. Scan a TMJ Open
image first and then scan a TMJ Close image. To prepare and position the
patient, follow the process outlined below.
For a child with a large head circumference, Man/Woman
mode may be selected instead of Child.
NOTE
Positioning the patient
1. Remove the standard chinrest and insert the sinus chinrest into the
equipment.
Disinfect the Chinrest using an alcohol-based cleaning
solution and wipe away all residues with dry cloth before
CAUTION
proceeding any further.
2. Insert the TMJ bite into the sinus chinrest.
PaX-i3D Smart(PHT-30LFO) User Manual

Hide quick links:

Tabla de contenido
loading

Este manual también es adecuado para:

Pht-30lfo

Tabla de contenido