PaX-i3D Smart(PHT-30LFO) User Manual
Aligning the Laser Beams
Be careful not to project the laser beams directly into the
patient's eyes as this could severely damage the patient's
WARNING
vision.
1. Mid-sagittal plane laser beam: Position the Mid-sagittal plane laser beam
on the center of the patient's face to prevent magnifications of the left or right
side in the final image.
2. Frankfurt plane laser beam: Position the patient's head so that the
Frankfurt plane is aligned with the Frankfurt plane laser beam.
3. Canine laser beam: Have the patient smile to properly position the canine
laser beam on the center of the patient's canine tooth.
3 Acquiring PANO Images
Frankfurt plane laser beam
UP button
Frankfurt plane laser beam
DOWN button
Canine laser beam lever
27