16. Error messages and troubleshooting

Error
message
The cuff does not inflate
sufficiently.
The cuff tube is not
connected correctly to the
device.
Er2
The cuff is not in the right
position on the arm.
Er3
Strong vibration during
measurement.
Er5
Measurement has failed
because you moved or spoke.
Er7
Measurement failed.
AFTER SALES SUPPORT
Possible cause
46 of 116
866-633-4660
Action
Attach the plug on the cuff tube
correctly to the device.
Tube or cuff leaky. Replace the
tube or cuff.
Put the cuff on correctly and
take the measurement again.
Put the cuff on correctly and
take the measurement again.
Keep still and do not speak
during measurement.
Keep still and do not speak
during measurement.
Put the cuff on correctly and
take the measurement again.