koban KPA-12 Instrucciones Operativas página 14

Medidor de abrazadera ac/dc
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Meter Description
1. Current clamp
2. Non-contact AC voltage indicator light
3. Clamp trigger
4. Rotary Function swith
5. Data Hold/ Back Light button
6. Relative button
7. LCD display
8. MODE select button
9. Range button
10. Hz% Hlod button
11. COM input jack
12. V ΩCAP TEMPHz jack
13. Battery Cover
1.
AC DC
AC (alternating current)
and DC (direct currrent)
2.
Minus sign
3.
8.8.8.8 4000 count (0 to 3999)
measurement reading
4.
AUTO
AutoRange mode
5.
REL
Relative mode
6.
Diode test mode
7.
•)))
Audible Continuity
8.
HOLD
Data Hold mode
℃,℉, μ,m,V,A,K,M,Ω, Units of measure list
9.
10. Hz %
Frequency/duty cycle test mode
Specifications
Function
Range & Resolution
AC Current
40.00 AAC
(50/60Hz)
400 .0AAC
40.00 ADC
DC Current
400 .0ADC
400.0 mVDC
4.000 VDC
DC Voltage
40.00 VDC
400.0 VDC
600.0 VDC
AC Voltage
400.0mVAC
www.grupotemper.com
KPA-12
AC/DC CLAMP METER
Ω
6
7
10
4
8
5
1
3
2
9
Accuracy (% of reading)
± (2.5 % + 8
)
digits
± (2.8 % + 5
)
digits
± (2.5 % + 5
)
digits
± (2.8 % + 5
)
digits
± (0.8% + 2 digits)
± (1.5% + 2digits)
± (2 % + 2
)
digits
± (1.5% + 30 digits)
3
KPA-12
AC/DC CLAMP METER
(50-400Hz)
4.000 VAC
40.00 VAC
400.0 VAC
600.0 VAC
400.0 Ω
4.000KΩ
40.00KΩ
Resistance
400.0KΩ
4.000MΩ
40.00MΩ
Capacitance
40.00nF
400.0nF
4.000μF
40.00μF
100.0μF
Frequency
10-10kHz
Sensitivity:100V(<50Hz);50V(50 to
400Hz;15V(401Hz to 10kHz)
Temp
-20.0 to 760.0°C
(type-K)
-4.0 to1400.0°F
(probe accuracy
not included)
Clamp size
Opening 1.2" (30mm) approx
Diode Test
Test current of 0.3mA typical; Open circuit voltage 1.5V DC
typical.
Continuity Check
Threshold <150Ω; Test current < 0.5mA
Low Battery Indication
"
" is displayed
Overrange Indication
"OL"" is displayed
Measurements Rate
2 per second, nominal
Input Impedance
10MΩ (VDC and VAC)
Display
4000 counts LCD
AC Current
50-60Hz (AAC)
AC Voltage bandwidth
50-400Hz (VAC)
o
Operating Temperature 41 to 104
F (5 to 40
o
Storage Temperature
-4 to 140
F (-20 to 60
Operating Humidity
Max 80% up to 87°F (31°C) decreasing linearly to 50% at 104°F(40°C)
Storage Humidity
<80%
Operating Altitude
7000ft. (2000meters) maximum.
Over voltage
Category III 600V
Battery
One 9V Battery
Auto OFF
approx. 30 minutes
Dimensions/Weight
197x70x40mm/183g
Safety
For indoor use and in accordance with Overvoltage Category II,
4
± (1.5% + 5 digits)
± (2.0% + 5
)
digits
± (1.0% + 4 digits)
± (1.5% + 2 digits)
± (2.5% + 3 digits)
± (3.5% + 5 digits)
±(4.0% reading + 20 digits)
±(3% reading + 5 digits)
±(4.0% reading + 10 digits)
±(1.5% reading + 2 digits)
± (3%rdg+5°C)
± (3%rdg+9°F)
o
C)
o
C)
www.grupotemper.com
Tabla de contenido
loading

Tabla de contenido