Process Display; Power Supply Connection - Endress+Hauser RIA 550 Manual De Utilización

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 76

Process display

Terminal layout
L
L for AC
N
N for AC
Ground connection
81
+ 24 V Loop power
82
0 V Loop power
83
+ 24 V Loop power
84
0 V Loop power
11
current sensor 50Ω
13
Measured signal Current, voltage, thermo-couple, RTD
14
Signal ground Current, voltage, thermo-couple, RTD
15
Power connection + RTD (2-/3-wire)
21
current sensor 50Ω
23
Measured signal Current, voltage, thermo-couple, RTD
24
Signal ground Current, voltage, thermo-couple, RTD
15
Power connection + RTD (2-/3-wire)
42
Connection relay 1
43
Switch contact
45
Connection relay 2
46
Switch contact
52
Connection relay 3
53
Switch contact
55
Connection relay 4
56
Switch contact
24
L+ for DC
N- for DC

4.2 Power supply connection

Before installing the unit please check that the power
supply used corresponds with that shown on the unit le-
gend plate.
Always connect the ground (earth) cable before connec-
ting any other cables.
When operating with a 90...253 VAC unit a power isola-
tor must be situated within easy reach of the unit. This
should also be fused with at least 10 A.
Connections/examples
In/output
Power supply
Loop power supply 1
Loop power supply 2
Input 1
Input 2
Relay output 1
Relay output 2
Relay output 3
Relay output 4
Tabla de contenido
loading

Tabla de contenido