PREPARATION BEFORE THE FIRST STERILIZATION :
CLEANING OF INSTRUMENTS
The instruments to be sterilized must be cleaned without any residue, such as fragments, blood, plaster,
resin, tampons, etc. These substances may cause damages to the objects in the basket or to the autoclave
itself.
FOLLOW THE DIRECTIONS GIVEN BELOVE:
-
clean the instruments after use to eliminate every residue
-
use, if possible, an ultrasound unit, with detergent solution with a basis of distilled water.
-
follow suggested instructions on use of the products for cleaning and lubrication of instruments
after using the ultrasound unit.
-
check all the directions given by the constructor to know
be put into the autoclave.
SETTING INTO THE BASKET OR INTO THE INSTRUMENTS HOLDER (optional)
-
Make sure that the instruments of different materials ( stainless steel, carbon
are separated.
-
in case of carbon steel instruments, put a paper napkin between
instrument, in order to avoid any contact of the two different materials.
-
put the instruments in the special holder inside its holes according to their diameter (Optional).
-
apply a piece of sterilizing indicating tape directly to the instruments or the basket (not on the
bottom).
-
use a biological indicating test to examine the lack of spores every
-
to have a perfect sterilization do not overload the basket with instruments.
-
empty containers must be left upsidedown to avoid any presence of water inside.
Autoclave GIMA H100 automatic
- 68 -
which
instruments can
steel, etc.)
the
basket and
six months.
the