This autoclave is an extremely simple and economic unit, it is equipped with a control panel with every
controls and required information for its proper use.
It has one sterilizing cycle which is optimized for a perfect and quick sterilization of the different materials
used in hospital and/or surgery.
The unit is made of a boiler with electric heating resistance, a door, a gasket, a timer to count sterilizing time
and a microswitch to check temperature.
Electric resistance is electronically checked to obtain a quick increase of water temperature and the steam
keeping at the set temperature.
Sterilizing cycle may be descripted in four phases:
1.
Loading of materials to be sterilized and locking of the door.
2.
Heating of the chamber, thermodynamic vacuum of air and next increase of temperature and steam
pressure to the set value.
3.
Beginning of sterilization for the set time (timer decrement).
4.
Manual steam exhaust or its condensation and bind on the bottom of sterilizing chamber.
Experience acquired in these years has brought to technological improvements for autoclave safety, expecially:
-
Safety lock , which prevents the opening of the door during the sterilization cycle; the door can
only be opened when pressure inside the chamber is levelled to the ambient one.
-
Safety valve for steam exhaust in case of damage with inside overpressure > 2.5 bar.
-
Safety thermostat which occurs if the inner temperature of the chamber rises over 135°C.
-
Thermodynamic vacuum.
Autoclave GIMA H100 automatic
4

DESCRIPTION