4. Operation
4.8. EC calibration options
Molar standards
147 µS/cm, 1413 µS/cm, 12.88 mS/cm and 111.8 mS/cm at 25 ˚C
(77 ˚F)
Demal standards
1049 µS/cm, 12.85 mS/cm and 111.31 mS/cm at 25 ˚C (77 ˚F)
NaCl standards
1014.9 µS/cm at 25 ˚C (77 ˚F)
Calibration to X value
Manual re-adjustment of the measured conductivity at any value
of the scale.
The EC readjustment is equivalent to calibration with only one
buffer.
10-04-2010
17:15
EC CALIBRATION
Channel 1
Molar standards
Calibration
Type of calibration
...
Start calibration
Cal. Frequency
Start calibration
Calibration
Type of calibration
Cal. Frequency
Cell constant introduction
Manual introduction of the cell constant. For example, when the
cell has been calibrated previously in a calibration laboratory.
Theoretical calibration
Selecting "theoretical calibration" the calibration data of the cell
in use are replaced by C=1.000 cm
This step is prior to obtaining a calibration certificate.
Calibration frequency
Select between 1 and 99 days.
The instrument by default is programmed for calibration every 15
days and indicates the remaining time to the new calibration.
Selecting 0 h automatically deactivates the re-calibration
warning.
Important note:
Once the calibration is re-programmed, it will be performed with
the new programming.
Calibration
EC CALIBRATION
Molar standards
Molar standards
Demal standards
NaCl standards
Use standard 147 uS
To a X value
Continue...
Data introduction
Calibration
EC CALIBRATION
Molar standards
Molar standards
Demal standards
Use
NaCl standards
standard
12.85 uS
To a X value
Continue...
Data introduction
Calibration
EC CALIBRATION
Molar standards
NaCl standard
Demal standards
Use
NaCl standards
Standard 1014.90 uS
To a X value
Continue...
Data introduction
Calibration
EC CALIBRATION
Molar standards
To a X value
Demal standards
NaCl standards
Use specific standard
To a X value
Continue...
Data introduction
Calibration
EC CALIBRATION
Molar standards
Data introduction
Demal standards
NaCl standards
C=1.000 cm-1
To a X value
Data introduction
Modify the data with
Accept with
The instrument goes to
MEASURE menu. The
introduced calibration
data will be used.
Calibration
EC CALIBRATION
Demal standards
Theoretical cal.
NaCl standards
To a X value
C=1.000 cm-1
Data introduction
Theoretical cal.
Continue...
Calibration
The instrument goes
Calibration frequency
to the option "EC
Automatic warning.
CALIBRATION".
1 day
00
hours
-1
.
Start of calibration with the selected
standards.
EC CALIBRATION
EC CALIBRATION
21,4˚C
* * *
83.1
uS/cm @25˚C
00:01
EC CALIBRATION
* * *
EC CALIBRATION
Adjust value
83.1
Accept with
EC CALIBRATION
The instrument goes
Adjust value
in MEASURE menu.
85.0
The theoretical cali-
bration will be used.
The instrument goes
to main menu.
21,4˚C
uS/cm @25˚C
00:05
21,4˚C
uS/cm @25˚C
00:07
uS/cm @25˚C
00:15
uS/cm @25˚C
00:15
11