7 Repeat the same procedure for the
lancing device (step 1 to step 6 ).
NOTE
• After the pre-cleaning and disinfection procedure, the
control solution should be tested to confirm that the meter
works properly before using the meter.
• Control solution tests should be performed using two
different levels of ReliOn™ Premier Control Solutions (Level 1
and/or Level 2).
• Verify that the test results are within the range printed on
the test strip vial.
• See page 28-30 for how to perform a control solution test.
NOTE
If any of the following deterioration signs appear after
pre-cleaning or disinfecting, please stop using the system and
contact Customer Service: 1-855-776-0662, 7 a.m. to 7 p.m.
Central time, 7 days a week.
• The inscriptions on the exterior of the meter or lancing
device have been removed.
• The color of the meter or lancing device has changed or faded.
• Cracks or roughness develop on the meter or lancing device.
• A part of the segment on the meter display does not flash.
• Control solution test results do not fall within the stated
range on the test strip vial.
54