19.5. Informes de la prueba de la pipeta

Ejemplo de impresión del informe estándar (KERN YKB-01N):
----------------------------Pipettes calibration-----------------------
Operator
Pipette
Serial no
Number of channels
Number of measurements
Operation with ISO 8655
Start date
End date
Water Temperature
Temperature
Humidity
Pressure
Z coefficient
------------------------------Tested volume: 1000 µl ------------------------
1
0.998 g
2
0.998 g
3
0.998 g
4
0.998 g
5
0.998 g
6
0.998 g
7
0.998 g
8
0.998 g
9
0.998 g
10
0.998 g
Average volume
Average [%]
Systematic error
Systematic error [%]
Permissible error
Random error
Random error [%]
Permissible error
Status
-----------------------------------------------------------------------------
Signature
....................................................................................................
AET/PET/ILT-BA-s-1631
Admin
RP-AF1000
GF202077
1
10
Ja
2015.03.05 11:12:30
2015.03.05 11:17:30
21°C
21°C
48 %
1007hPa
1.00283
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
1000.82389 µl
100.08 %
0.82389 µl
0.082389 %
± 16 µl
0 µ
0 %
± 6 µl
positive
205