Change The Measurement Units; Use A Sample Id; Select The Data Output; Change The Date And Time - Hach sensION+ MM340 Manual De Usuario

Ocultar thumbs Ver también para sensION+ MM340:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 38
1. From the main menu use the
MEASURE. Confirm.
2. Use the
to change the following settings. Confirm every entry.
Option
Description
Resolution
Select the resolution: 1, 0.1, 0.01 (default) or 0.001
Measure
Stability—select By stability Criterion: Fast (variation <
0.02 pH in 6 s), Standard (variation < 0.01 pH in 6 s) or Strict
(variation < 0.002 pH in 6 s). In continuous—enter the time
interval for the In continuous Acquis. interval (data storage or
printing data). By time—enter the time interval for data storage
or printing data.
Display mV
Display mV—select YES or NO to show mV.
Limits
Limits—select YES or NO. YES: Enter the upper and lower
limits. An acoustic warning appears when the measurement is
out of limit. The report output shows an A next to the
measured value when the measurement was out of limit.
Isopotential Isopotential—change the Isopotential pH value in Data
introduction. Select Calculate to calibrate the probe again.
3. Push
to start the measurement.
Note: If the measurement is not stabilizing after 120 seconds, the meter turns
automatically into the continuous measurement mode.
Section 7 Advanced operation

7.1 Change the measurement units

The measurement units can be changed individually for each channel.
1. From the main menu use the
2. Use the
or
to select Measurement units and confirm.
3. Select Channel 1 or Channel 2 and confirm.
4. Select ORP (mV), pH or ISE and confirm.
or
and and to select
or
to select SYSTEM. Confirm.

7.2 Use a sample ID

The sample ID tag is used to associate readings with a particular sample
location. If assigned, stored data will include this ID.
1. From the main menu use the
2. Use the
or
to select Sample ID and confirm.
3. Use the
or
to select
Option
Description
Automatic A consecutive number will be automatically assigned to every
sample.
Manual
A keyboard or a barcode scanner is required to enter the
sample ID name before taking a measurement (maximum
15 characters).

7.3 Select the data output

Data can be stored or transferred to a printer or to a PC.
1. From the main menu use the
2. Use the
or
to select Data Output and confirm.
3. Use the
or
to select
Option
Description
Deactivated
Select Deactivated if no printer or PC is connected.
For Printer
Select Dot matrix printer or Thermal printer.
For Computer Select Terminal, LabCom or LabCom Easy. The LabCom
Software controls several modules, pH and conductivity
meters, automatic burettes, Samplers and so on from a
computer. The LabCom Easy software gets measurements
and calibration data from the meters

7.4 Change the date and time

The date and time can be changed from the Date / Time menu.
or
to select SYSTEM. Confirm.
or
to select SYSTEM. Confirm.
English 11