Error Messages - Agilent Technologies Turbo-V 301-AG Manual De Istrucciones

Tabla de contenido

Error Messages

For a certain type of failure, the controller will self-diagnose the
error and the messages described in the following table are
displayed.
Tab. 1
MESSAGE
CHECK
CONNECTION
TO PUMP
WAITING
INTERLOCK
PUMP
OVERTEMP.
CONTROLLER
OVERTEMP.
Turbo-V 301-AG Rack Controller User Manual / 87-900-982-01
DESCRIPTION
Wrong connection between the
pump and the controller.
Or
The pump's temperature is below 0
°C.
The interlock signal of P1
connector is activated by an
interruption of the link between pin
3 and 8 of J1 connector, or
because the external interlock
signal is open.
The upper bearing/pump
temperature exceeds 60 °C.
The controller environment
temperature exceeds 70 °C.
Or
The controller's radiator
temperature is above 60 °C.
Instructions for Use

Error Messages

REPAIR ACTION
Check connection between
controller and pump.
Press the START push-button
twice to start the pump.
Reset the short circuit
between pin 3 and pin 8 of P1
connector, or close the
external interlock signal.
Wait until the temperature
decrease below threshold
value.
Press the START push-button
twice to start the pump.
Wait until the temperature
decrease below threshold
value.
Press the START push-button
twice to start the pump.
(F)
16
205/294
Tabla de contenido
loading

Este manual también es adecuado para:

969-8991969-8992969-8993

Tabla de contenido