Performance
Flow rate
Timing
Positive Pressure
Safety Relief
Valve
Gauge Accuracy
Adult
Pediatric
Neonatal
Physical
Dimensions
Weight
Battery
Lithium
full scale deflection
1
full range at 22°C
2
not adjustable, without fittings at full increase
3
Timing is preset at factory using a source vacuum level of 500 mmHg (66.7 kPa). Timing may vary in relation
4
to the source vacuum pressure.

ENVIRONMENTAL SPECIFICATIONS

Operating Temperature Range
Storage and Transport Temperature Range
Operating Atmospheric Pressure Range
Storage and Transport Atmospheric Pressure Range
Operating and Storage Relative Humidity
Ingress Protection Rating
8700-0009-000 (Rev. 16) 10/2020

Specifications

Intermittent
Continuous
15 seconds ON, 8 seconds OFF (± 3 seconds)
Located in the vacuum supply line to prevent pressurization of the patient
connection by failed injector vacuum (Venturi) units, or inadvertent cross
connection to pressured gases.
0-200 mmHg (0-26.7 kPa)
0-160 mmHg (0-21.3 kPa)
0-160 mmHg (0-21.3 kPa)
0-100 mmHg (0-13.3 kPa)
6.5"H x 2.8"W x 4.8"D
(16.5 cm x 7.1 cm x 12.2 cm)
20 oz (567 grams)
Two 2/3 AA batteries, 3.6V, 1.6 Ah, lithium
Adult
Pediatric
8 L/min
8 L/min
80 L/min
80 L/min
3
0-16 L/min (Preset at 8 L/min)
Can be adjustable
Starts in the ON cycle
Analog
1
±5% Full-scale
±5% Full-scale
±5% Full-scale
-
50 to 104ºF (10 to 40ºC)
-13 to 158ºF (-25 to 70ºC)
80 to 106 kPa
50 to 106 kPa
5 to 95% RH (Non-condensing)
IP 20
Neonatal
8 L/min
80 L/min
3
3
4
Digital
2
±1% Full-scale
±1% Full-scale
-
±1% Full-scale
5