Cleaning and Care

Clean the rotor and the accessories regularly and in
case of need. This is indicated both for reasons of hy-
giene and to prevent corrosion due to contamination
sticking to the instrument and its accessories.
Clean them with mild agents of pH values ranging from
6 to 8.
Immediately after cleaning, dry the aluminum parts or
put them into a warm-air dryer at a temperature not
exceeding 50° C.
Disinfection
If a capillary tube containing infectious material leaks,
during a centrifugation run, you have to disinfect im-
mediately the rotor.
Infectious material could enter the
centrifuge if spills or tube breakage
occur.
Danger of infection may occur upon
contact!
Take appropriate protective measures
for personnel!
Mind the permissible filling volumes
and loading limits for the tubes!
Broken glass capillaries have sharp
edges and pose a risk of injury.
In case of contamination the operator
has to make sure, that no further per-
sons are jeopardized!
Contaminated parts have to be decon-
taminated immediately.
If required further protective measures
have to be initiated
.