9. OUTPUT CONNECTORS
AC Output:
An AC signal corresponding to the frequency-weighted signal is available at this
connector.
Output voltage: 1Vrms±50mVrms (scale upper limit)
Output impedance: approx. 5kΩ
Load impedance:≧1MΩ
The output voltage when the meter is in calibration mode
(-16dB from scale upper limit, 1000Hz sine wave) is 158.5mVrms.
DC Output:
A level-converted DC signal generated by RMS detection and logarithmic
compression is available at this connector. The signal reflects the frequency
and time weighting settings of the instrument.
Output voltage: 10mV±0.1mV/dB
Output impedance: approx. 5kΩ
Load impedance:≧1MΩ
The output voltage when the meter is reading 114dB is nominally 1.14V DC.
10. SOFTWARE INSTALLATION and OPERATION
For the detailed instruction, please refer to the content of attached CD-ROM,
which has the complete instruction of software operation and relevant
information.
09/15/17 Version No. 00
Datalogger Sound Level Meter / EN