Instrument Upgrade; Returns; Disposal; Technical Data - Tintometer Lovibond SD 400 Oxi L Instrucciones

Ocultar thumbs Ver también para Lovibond SD 400 Oxi L:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 66
Large measurement value
fluctuations

11. Instrument upgrade

In general, it is not necessary to perform an upgrade. If software updates are available, they
can be downloaded from our homepage: www.lovibond.com.

12. Returns

All instruments which are sent back to the manufacturer must be free from meas-
urement substance remains and other harmful substances. Measurement substance
remains on the housing or on the sensor can endanger persons or the environment.
Use suitable transport packaging for the return of the instrument, especially if it is still a
functioning instrument. Make sure that the instrument is protected with sufficient insulating
material in the packaging.

13. Disposal

Dispose of depleted batteries at a collection centre designated for this purpose.
The instrument may not be disposed of with the household waste. If the instrument needs
to be disposed of, send it directly to us (with sufficient postage paid). We shall dispose of the
equipment properly and in an environmentally friendly manner.

14. Technical data

Sensor
Oxygen, dissolved
- Measurement range
- Precision
- Resolution
Temperature
- Measurement range
- Precision
- Resolution
Sensor membrane
Dimensions
Device
Sensor
SD_400 Oxi L_1 01/2016
a) Temperature fluctuations
b) Sensor cap dirty or de-
fective
c) Sensor not sufficiently im-
mersed in the sample
d) Sensor damaged
e) External electronic fault
Optical (measurement of the luminescence life)
0 – 50 mg/l or 0 – 500 % air saturation
0 – 20 mg/l or 0 – 200 %: ± 1 % of meas. or
±0.1 mg/l (higher value applies)
>200 % or >20 mg/l: ± 10 % of meas.
0.1 %, 0.01 mg/l
-5 – 50 °C (storage and operating temperature)
± 0.2 %
0.1 °C
Plastic
approx. 162 x 97 x 50 mm (L x W x H)
approx. 287 length, ø 33 mm incl. storage flask
a) Keep temperature con-
stant
b) Clean or replace sensor
membrane
c) Immerse sensor sufficiently
deep in the sample (at least
3 cm)
d) Replace sensor
e) Switch off or remove ex-
ternal sources of interference
31
Tabla de contenido
loading

Tabla de contenido