Protocolo de rescate en modo DEA (Reanálisis automático
You pressed the ON/OFF button to
turn ON the Paramedic CU-ER3
CU-ER3 detects/decides:
AED electrode pads
Voice and text prompts:
DO NOT TOUCH THE PATIENT
Voice and text prompts:
ANALYZING HEART RHYTHM
CU-ER3 detects/decides:
Voice and text prompts:
SHOCK ADVISED
Voice and text prompts:
STAND CLEAR
Text prompt, flashing:
CHARGING
Speaker playback:
SIREN SOUND
At the end of charging:
Text prompt:
CHARGING COMPLETE
Beeper:
A burst of two beeps
Voice and text prompts:
PRESS THE FLASHING RED
BUTTON, NOW
While the Paramedic CU-ER3 is
waiting for the SHOCK button to
be pressed:
Beeper:
Continuous tone
Shock button backlight:
Flashing red light
CU-ER3 detects/decides:
Within 15 seconds, SHOCK
PRESSED
Voice and text prompts:
SHOCK DELIVERED
APÉNDICE B
Start
Paramedic
Voice and text prompts:
NO
ATTACH PADS
attached?
YES
Paramedic
ECG Shockable?
YES
Paramedic
NOT PRESSED
button was :
Voice and text prompts:
THE SHOCK BUTTON WAS
NOT PRESSED
Voice and text prompts:
BEGIN CPR, NOW
Voice and text prompts:
PUSH THE CHEST DOWN, FAST,
TWO INCHES
Text prompt:
CYCLE NO.
COMPRESSION NO.
Beeper:
Beat Sound at 100 beats per minute
Voice and text prompts:
GIVE TWO BREATHS
BREATH, BREATH
This CPR sequence is done for 5 cycles
CU-ER3 Operator's Manual _ver 1.00
NO
Voice and text prompts:
NO SHOCK ADVISED
Voice and text prompts:
CHECK PULSE
Timer
10-second interval for
checking pulse
Voice and text prompts:
IF NO PULSE, BEGIN CPR
)
130