Operating instructions Midwest
Rhino XE
67 16 786 D3758.201.01.01.02
Procedure
ü
The instrument is removed [→ 20].
ü
The black nozzle insert is in the spray can tube.
1.
Detach the motor from the supply hose [→ 19].
2.
Set the drill direction to F.
3.
Insert the nozzle of the spray can in the driving air tube. Hold
the motor with a cloth.
4.
Spray into the motor for 1-2 seconds.
IMPORTANT: Hold the spray can upright.
5.
Wipe any spray that comes out with a disinfection cloth.
6.
Repeat the process until the spray is clear.
7 Reprocessing
29