Output check
The output check includes a voltage check under no load through all channels; additionally it
checks for short circuits, this is to ensure the instrument correctly functions.
1. Voltage output check
(1) Within no load, turn on the power, and make sure the current setting of all channels is not
zero.
(2) Turn on CH1/CH2 and set the instrument to constant voltage mode. Check whether the
voltage value could be changed from 0V to 32V.
2. Current output check
(1) Turn on the power, and make sure the current setting of all channels are not zero.
(2) Use an insulated wire to connect the positive and negative terminal of CH1/CH2.
(3) Turn on CH1/CH2 and set instrument to constant current mode. Check whether the
current value could be changed from 0A to 3.2A.
14