6. To begin calibrating, carefully press the Probe Tip onto
the Calibration Disk perpendicularly as shown in Fig.
1.13. Ensure that the Probe Tip remains fully depressed
during this process, to achieve consistent pressure on
the tip.
Fig. 1.13
7. During calibration, the Lights Simulant, Moissanite
and Diamond Indicator lights will each gradually light up,
accompanied by rhythmic beeping, with Diamond LED
lights indicating the ending phase of the calibration, as
shown in Fig. 1.14.
*Note : Do not lift up the tip from the Calibration Disk.
Continue to hold the position.
Fig. 1.14
8. When calibration is complete, the Indicator Lights
will be fully lit and a long beep will sound, followed by
a unique dual beep. This indicates that the tester is fully
calibrated and ready to use.
13
Diamond
Moissanite
Simulant
Metal
Low Batt
Starting of
calibration
Diamond
Moissanite
Simulant
Metal
Low Batt
Calibration
Nearing Completion