Perform the Product electrical
Making any changes to this
device is forbidden.
Interrupt the power supply before
doing any maintenance jobs.
Check Product integrity.
N
1
Before using
2
Before using
3
Once a year
4
Once a year
MU_023
Rev.2
check.
Period
28/10/2019
OPERATION AND
MAINTENANCE MANUAL
7.3 Periodical checks to be performed on
the Product
At the time of start up and after each maintenance job, perform
electrical tests and jobs indicated in the IEC 62353 standard.
7.4 Routine maintenance
Make sure there are no pieces or fragments of paint that could
become detached and fall within the operating field. If there are any,
remove them manually.
Make sure the light source protection screens are not damaged. If
they are, contact the Customer Service.
Check all the Product joints and make sure there are no noises or
squeaks. If there are, lubricate the clutches involved with suitable
grease for industrial use at a service temperature between -30°C
and + 120°C, type OKS 470 or with similar properties.
If the Product fails to maintain a regular position, adjust the clutches
as indicated at points 7.1 and 7.2 (arm and clutch adjustment).