It may be necessary to prime the system by squeezing
9 9
gently on bottle. This will cause the tubing to be filled
with supplement and prevent the infant from swallow-
ing air.
After use, wind tubing
1 1 0 0
around attachment ring
and close system with
cover.
CLEANING & STERILIZATION
1. Hospital Sterilization
AUTO-
CLAVE
GAS
All disassembled parts can be auto-
claved to a maximum of 272°F for 3
minutes at 29 PSIG or 250°F for 15
minutes at 15 PSIG.
To avoid damage during sterilization
-
Sterilize parts without applying
external pressure on parts.
If gas sterilization is used, residual test-
ing is recommended to determine the
proper length of aeration.
7