Troubleshooting The Trainer - Laerdal HeartStart Trainer Instrucciones De Utilizacion

Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 19
The following table is provided to help you troubleshoot potential problems.
BEHAVIOR
Trainer says ANALYZING
before pads are applied to
manikin.
Trainer does not say ANALY-
ZING when pads
are applied to the manikin
adapter.
Trainer says ANALYZING
while yellow liner is still
attached to pad.
The training pads cartridge
is difficult to install in or
release from the Trainer.
Trainer's voice instructions
high-pitched and too rapid.
16

Troubleshooting the Trainer

POSSIBLE CAUSE
• User is touching the adhesive gel and
activating the sense circuitry of the Trainer.
• Liner is not properly separating pads.
• Wires are twisted.
• Pads placement is incorrect.
• Pads are old.
• Replaceable pads are not properly
connected to cartridge cable.
• Manikin adapter is broken.
Trainer cannot sense yellow liner.
The resealable film cover may be
improperly aligned on the cartridge.
One of the batteries is installed backwards.
RECOMMENDED ACTION
• Avoid touching the gel on the pads.
• Be sure the pads are pressed firmly
to the liner and only touching each
other through the liner hole.
• Untwist the wires.
• Check for proper pad placement
• Replace training pads.
• Check latch to be sure the
replaceable pads are correctly
attached to the cartridge cable.
• Replace manikin adapter.
Remove yellow liner.
NOTE:This cannot happen when using
the HS1 Defibrillator.The defibrillator will
not analyze heart rhythm when the
yellow liner is attached to the pads.
It senses that the pads are not attached
to a patient and advises the user to re-
move the yellow liner.
Make sure the resealable film cover
is lined up properly when repacking
the cartridge.
Be sure to install batteries according
to the diagram in the Trainer's battery
compartment.