Functional Test; Preparing Tropic 2.0; Nebulizing Directly From Nebulizer Chamber - Krömker TROPIC 2.0 Instrucciones De Uso

Nebulizador ultrasónico
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 90
EN
6.3

Functional test

Check the functioning of the device prior to every use and after every reprocessing procedure.
Only device parts in faultless condition can ensure the functionality of the ultrasonic nebulizer. For this reason, the
device parts must be subjected to a thorough check and inspection prior to any assembly.
Check and inspect:
if the mains cable is undamaged,
-
if the device parts made from plastic and rubber (e.g. hose, nebulizer chamber) are in faultless condition and
-
do not show signs of age-related damage,
if the device was properly cleaned and there is no residue or soiling.
-
6.4

Preparing TROPIC 2.0

6.4.1

Nebulizing directly from nebulizer chamber

96200-xxx_00 GA USV Tropic 2.0_20160229.docx
Connect the power cable (D) to the device socket and to a
-
grounded mains power socket.
Inserting the nebulizer chamber
Fill the nebulizer chamber to the max. mark [2] with nebulizing
fluid (2).
Plug the nebulizing chamber (2) into the holder of the TROPIC 2.0
[1]. Observe the preferred position.
Press the lid onto the nebulizer chamber
Mounting the bacterial filter
Connect the bacterial filter [1] to the fan connection [2] on the
housing
Observe the flow direction according to the manufacturer
specifications.
- 50 -
Tabla de contenido
loading

Tabla de contenido