Electromagnetic compatibility (EMC)

Invia Liberty is EMC-tested in conformity with the requirements of IEC
60601-1-2:2007 and IEC 60601-1-2:2014 4th Edition according clause 7
and 8.9.
Electromagnetic compatibility ( EMC, IEC 60601-1-2:2007 3rd Edition and
IEC 60601-1-2:2014 4th Edition, Table 1 )

Electromagnetic emissions

The Invia Liberty pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the Invia Liberty pump should assure that it is used in such environment.
Emission tests
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
WARNING
The Invia Liberty pump should not be used adjacent to or stacked with other equipment.
If adjacent or stacked use is necessary, the Invia Liberty pump should be observed to verify normal operation
in the configuration in which it will be used.
WARNING
Invia Liberty is a medical device that requires special safety precautions and must be installed and placed in
operation in accordance with the attached EMC information.
WARNING
Do not use other accessories than those specified or sold by the manufacturer as replacement parts for
internal components as it may result in increased emissions or decreased immunity of the Invia Liberty pump.
WARNING
EMC
HF (high-frequency) surgical equipment, radio networks or the like can influence the operation of the device
and may not be operated in combination with the Invia Liberty pump.
WARNING
Portable RF communications equipment (including peripherals such as antenna cables and external
antennas) should be used no closer than 30 cm (12 inches) to any part of the Invia Motion. Otherwise,
degradation of the performance of this equipment could result.
Compliance
Electromagnetic environment – guidance
Group 1
The Invia Liberty pump uses RF energy only for it's
internal function. Therefore, its RF emissions are very
low and are not likely to cause any interference with
nearby electronic equipment.
Class B
The Invia Liberty pump is suitable for use in all establish-
ments, including domestic establishments and those
directly connected to the public low-voltage power sup-
Class A
ply network that supplies buildings used for domestic
purposes.
Complies