OPERATOR'S MANUAL
60
62
78
®
mini GAC
11001-1464 Rev C
ERROR 56
Probable Cause
Error 56 displays when the measured average tilt angle exceeds
10 degrees after an Empty Cell and/or a Full Cell test phase.
Corrective Action
1. Hold mini GAC level while performing test or
2. Place on level surface to perform test.
3. Press the Enter button and perform a new test.
ERROR 60
Probable Cause
Language files are not found in device and can only be loaded
by a DICKEY-john technician.
Corrective Action
1. Power off device. Unit is not operational and should be
returned to DICKEY-john for service.
ERROR 62
Probable Cause
The language selected is not loaded on the instrument and the
first language is automatically selected.
Corrective Action
1. Press the Enter key to return to the Main Menu. Select an
available language, if desired.
2. Return instrument to DICKEY-john to load a specific
language, if desired.
ERROR 78
Probable Cause
A normalization file required for proper operation is missing.
CORRECTIVE ACTION
1. Cycle power to see if error returns. If the error appears
again, return the instrument to DICKEY-john for service.
ERROR CODES / 31