28 / ERROR CODES
ERROR 12
Probable Cause
The measured temperature is above maximum. The cell
temperature is measured on startup and Error 12 displays if the
temperature is above +158 degrees F (70 degrees C).
Corrective Action
1. Pressing the Enter button causes the cell temperature
measurement to repeat and operation will proceed if the cell
temperature is below the maximum, otherwise Error 12
displays again.
2. If cell temperature is still above maximum, wait to test until
unit temperature is less than 158 degrees F (70 degrees C).
ERROR 20
Probable Cause
Error 20 displays if the measured product moisture is below the
lower limit of the production calibration. NOTE: Results may not be
accurate. Sample could be too dry to be measured accurately.
Corrective Action
1. Press Enter to show measured results.
ERROR 22
Probable Cause
Error 22 displays if the measured product moisture is above the
upper limit of the product calibration. NOTE: Results may not be
accurate. Sample could be too wet to be measured accurately.
Corrective Action
1. Press Enter to show measured results.
ERROR 28
Probable Cause
The calibration checksum of a newly selected product is incorrect.
Corrective Action
1. Press the Enter button to return to the Main Menu. Operation
can continue with an invalid calibration checksum. It is
mini GAC
11001-1464 Rev C
®