OPERATOR'S MANUAL
10
®
mini GAC
11001-1464 Rev C

ERROR CODES

Error codes display when an abnormal event occurs. An error
can be acknowledged by pressing the Enter button to return to
normal operation.
IMPORTANT:
Contact DICKEY-john Technical Support or a
distributor for any Error Codes that display
and are not listed.
LOW BATTERY VOLTAGE
Low Battery voltage displays when the battery voltage falls
below 6.4V but is above 6.2V. When the battery voltage is low,
the unit is still capable of making accurate measurements, but
the battery should be replaced soon. Several warnings will
display before the dead battery screen appears.
DEAD BATTERY WARNING
Displays when the battery voltage falls below 6.2V. When the
battery voltage is lower than 6.2V, the unit is not capable of
making accurate measurements and the unit will no longer
operate other than to power off.
ERROR 10
Probable Cause
The measured temperature is below minimum. The unit
temperature is measured on start up and displays if the
temperature is below 32 degrees F (0 degrees C).
Corrective Action
1. Pressing the Enter button causes the unit temperature
measurement to repeat and operation will proceed if the
cell temperature is above the minimum, otherwise Error 10
displays again.
2. If unit temperature is still below the minimum, wait to test
until the temperature is greater than 32 degrees F (0
degrees C).
ERROR CODES / 27